Flexible Stem Trabecular Prostheses

  • Share this article on Facebook
  • Share this article on Twitter
  • Share this article on Linkedin

Proper identification of isostatic lines and isorigidity regions for the sintered metal trabecular lattices distribution between the proximal and distal part of the stem has been carried out. The faithful models enables us to properly take into account not-isotropic properties of the femur proximal end, while clarifying the critical mechanical role of the trabecular bone that should be taken into account to design new innovative prosthetic system.

mediaimage
  1. Introduction

 

The validation of the clinical efficiency and the estimate the long-term reliability of prosthetic restorative systems need the appropriate understanding of the physical variables that influence the biomechanical behavior of the material for biomedical advanced applications.

The tool of the Finite Element Analysis (FEA) is allowing biomaterials researchers to attain comprehensive evaluation of the biologic and mechanical behaviors of advanced restorative systems,Flexible Stem Trabecular Prostheses Articles even in the case of not homogeneous systems.

If validated by proper experimental procedures, the FEA turns useful in the optimization of the restorative design criteria and in the choice of the materials to be used. Moreover, this method allows the estimate of the location of fractures under given loading circumstances (Aversa et al., 2016 a-o, 2017 a-e; Mirsayar et al., 2016-2017).

New fabrication processes based on additive manufacturing technologies and studies on biomechanics and biomimetics (Annnunziata et al 2006, Apicella et al 2010, Aversa et al 2009) could enable the set-up of new design criteria for human prostheses. The Authors have taken up these studies to gathering the unexploited potential of such advanced materials and design technologies by developing biofidel Finite Element models able to correctly mimic the femur biomechanical behavior (see fig 1).

Although human femur has an internal structure that, through the evolutionary optimization of the mass and disposition of the cortical and trabecular bone types (Tamar Gottesman and Zvi Hashin 1980, Oh and Harris 1976), is able to bear high external stresses (Ashman et al 1984, Dalstyra 1993).

The progressive physiological bone mass loss occurring in elderly people or in presence of prolonged inactivity, leads to the bone toughness weakening and to a reduction of its ability to resist to mechanical shocks.

This reduced resilience is the main cause of elder people pertrochanteric femur fracture (Ashman and Rho 1988, Burnstein  et al 1976, Carter and Hayes 1977).

The reduction of these fractures and recovery of the hip joint functionality needs the application of special prosthesis.

Presently, total hip replacements performed on patients younger than 65, and therefore with a higher life expectation time needing a longer durability of the orthopedic implant, do not endure for more than 15 years.

The orthopedic prostheses used today are made of metal alloys, plastics and ceramics materials with well defined properties and characteristics.

In particular, due to their high biocompatibility and high mechanical strength, some metal alloys based on Titanium or Cobalt Chrome are commonly used and they represent almost all of the prosthetic components.

These traditional prosthetic products are made from assessed technologies such as metal forgings or melting followed by mechanical machining.

These processes do not allow the optimal design solutions for a better biomechanical bone integration.

The application of other scaffolding material on the surfaces to create a more osteoinductive and osteoconducive substrate for a healthy bone ingrowth has to be necessarily undertaken (Schiraldi et al 2004, Gramazzini et al. 2016).

However, these processes, even if very expensive, do not provide a sufficient structural support and, often, they are not mechanically adequate.

Full Metal implants, on the other hand, can allow a quick and satisfactory structural restoration of functional mobility while maintaining acceptable running costs.

However, other than the high invasiveness, these implants do not represent the ideal solution, especially for younger patients who have high life expectation, both in terms of duration and biomechanical osteointegration.

This first-generation of prostheses, in fact, do not technically represent the optimal structure for a good osseo and bio-mechanical integration with the bone living tissues of the area surrounding the implant.

Comparing a metal prosthesis with a human bone at a radiographic examination it is evident the difference in density and mass that can induce not healthy bone growth (Apicella and Aversa 2015).

Frost (1987) reports that bone adaptive properties depend on the ranges of physiological strains to which bone is subjected. Implantation of the prosthesis could than favor reabsorption processes leading to more and a more insecure and unstable condition that, until the prosthesis shows signs of weakness, will move from its location producing serious consequences and pain to the patient. In more severe cases the implant breaks down, making it necessary for its immediate replacement.

Biomechanics and biomimetics: a way to foster advanced materials and technologies potentials

The implants are then expected to serve for much longer period without failure or surgical revision.

The design and improvement of correct combinations of materials and prosthetic systems showing higher endurance and improved biocompatibility becomes mandatory.

The prosthetic implant replacing the resected bone should expected to ensure an “equivalent stiffness” (combination of material elastic modulus and prosthesis shape) matching that of the missing part of the bone and that of the residual osseous region where it is implanted.

Due to its orthotropicity and differently density of the structure, the elastic modulus of the bone may vary from 4 to 20 GPa, which depends on the bone type and direction of loading.

The metal and ceramic implant materials adopted today, which are characterized by stiffnesses higher than those of bone where they are implanted, strongly alters the stresses physiological local distribution of the deformation with the unwanted bone reabsorption around the implant and, consequently, with an increase of the risk of implant failure.

This biomechanical mismatching leads to the lowering of the biological strain level needed for healthy bone growth (Frost, 1994) and to the loss of the bone mass due to this low level of loading know as stress shielding.

A customized material with improved combination of the needed strength and stiffness that match that of the bone has to be used for biomechanical integration of the implant.

This higher biomechanical compatibility will avoid implant loosening and will increase the its service lifespan avoiding additional surgery for revision and permitting a good bio-integration (bone growth).

Innovative biomimetic materials for tissue engineering that are based on hydrophilic polymers have been developed by our research group and have shown attractive physical, biological and mechanical properties for applications in the biomedical field (Schiraldi et al 2004).

Highly biocompatible novel hybrid scaffolding materials based on fumed silica and hydrophilic poly-(hydroxyl-ethyl-methacrylate) (pHEMA) to be used in concurrence to metal prostheses have been developed by the Authors.

The structural metal scaffold of the new prostheses will be made of Titanium alloys using the additive technology based on the principle of melting thin layers of Titanium powder (about 50 microns) one over the other until the achievement of the desired component is completed (see lower right of figure 9).

Then the nanostructured and osteoconductive biomimetic material developed in our previous studies (Schiraldi et al 2004) can cover the Titanium structural prosthetic scaffold.

These biological hybrid prostheses, which are made using synthetic materials able to induce the growth of biological networks and a metal structural scaffold, may favor the birth of new classes of hybrids orthopedic solutions in the medical field.

The new hybrid bio-prosthesis could drastically reduce the stress shielding phenomena while providing an advantageous improvement of prosthesis lifetime compared to traditional solutions.

The recovery of optimal joint functionality will lead to the improvement of the quality of life of the patient, who perceives significantly abating the risk of new surgical operation.

The requirement of envisaging the potential structural alteration that can be induces by the inadequate use of not biologically compatible prostheses in the structure and morphology of the bone, has forced our studies in the direction of the assessment of faithful models of the femur that could account for real bone type distribution and to its orthotropic behavior (Aversa et al 2016).

The development of faithful models has already enabled us to prepare prostheses that could restore the natural physiological stresses and strains patterns in the bone (Apicella 2010, Gramanzini et al. 2016, Perillo et al 2010, Sorrentino et al 2007 and 2009).

In this work a femur FEM model, which has been developed in a previous work to correctly represent structural behavior of the femur head (Aversa et al 2016), has been analyzed for the stress and strains distribution over the entire bone stem and head, and it has been modified to account for femur head resection and substitution with a Titanium hip joint prosthesis.

A comparison between the biological stress and strain distributions in the femur models of the sound and prothesized femurs could help in the understanding of the correct design procedures needed to design new innovative biomimetic prostheses.

 

 

  1. Materials and Methods

 

Medical Image Segmentation has been derived from CT using the Mimics software (Materialise, Belgium) to process a patient medical image.

As reported in Figure 2, processing of CT resulted in a highly accurate 3D solid model of the patient pelvis and femur anatomy.

New prosthetic engineering applications through a combined use of Mimics and 3-Matic (Materialise, Belgium) software’s could be derived from the study of the biomechanics of these bone districts.

A 3D solid and Finite Element Models (FEM) have been developed in a previous work to simulate the external and internal morphology of the femur as already done for other complex bone structures accounting for the orientation and densities of the head trabecular systems (Aversa R. et al. 2016, Aversa R, et al. 2009, Apicella, D. 2010, Beaupre G.S. and Hayes W.C. 1985, Reilly D. and Burstein A. H. 1974, Reilly D. T. and Burnestain A. H. 1975). The procedure is illustrated in the Figures from 3 to 6.

The external geometry of femur and portion of the pelvis have been reconstructed by generating a three-dimensional volume that interpolates the CT scans (Figure 3).

The results were then imported in the 3Matic software for surface and solid meshing optimization, Finite element model preparation and material properties definition (Aversa et al 2016, Apicella D. et al 2010, Apicella D. et al. 2015).

The results of tethraedric element material distribution are shown in fig.4 (the yellows correspond to different stiffnesses cortical bone mechanical properties while greens are relative the trabecular bone of different densities) .

The same procedure was applied to the preparation of the Finite Element Analysis (FEA) model of the femur with the resection of the head, which has been virtually done at the fracture plane reported in figure 1, and for the Titanium traditional prosthesis.

Internal modeling of the resected femur and the of the meta prosthesis have been realized by defining three-dimensional internal tethraedric meshing distribution and size optimization, as indicated in the upper part of Fig. 5 and in fig. 6, respectively.

The solid mesh elements of the resected femur have been sequentially associated to the patient bone densities according to the Hounsfield (HU) scale.

This scale linearly quantifies the X-rays attenuation coefficients in the tissues, assigning, by using the Mimics software, the proper elastic moduli to the corresponding elements of the FEM model (lower part of fig. 5).

The meshed solid models of the resected femur and of the Ti prosthesis were assembled accounting to the correct positioning of the implant using the Mimics software (figure 7).

New remeshing of the tethraedric elements were run in 3Matic (figure 8) searching for the node congruency at the bone-implant interfaces.

The mechanical properties have been assigned considering the cortical and trabecular bone characteristics.

In particular, the systems have been considered as isotropic materials and the mechanical properties have been coupled with each single tethraedric element characterized by equivalent Hounsfield (HU) densities scale.

The elastic and shear moduli of the trabecular bone were derived as a fraction of those of the cortical bone cortical bone with a direct proportionality with the porosity measured by the axial tomography. In this scale, the value of the fat is about -110, the muscles around 40, the trabecular bone stay between 100 and 300 while the cortical one covers the values between the high density cortical bone (300) and about 2000, that matched in our bone systems to Elastic moduli ranging from 0,87 to 15.0 GPa.

The solid mesh elements of the trabecular-oriented material have been assigned by operating on the internal structure accounting for the real bone trabecular morphology that we found in the proximal end, (fig. 6) as indicated in Aversa et al (2016).

Variable rigidity Ti prostheses Design

In order to avoid excessive stress shielding effects on the resected femur diaphysis (shaft), trabecular hip joint prosthesis has been properly designed according to the characteristic and specific rigidity of each diaphysis section interested by the prothesization.

The mechanistic model of the hip proximal epiphysis proposed by Kummer (1986) has been transferred to the hip joint prosthesis design where the presence of isostatic lines that characterizes the oriented trabecular systems, is reported in Figure 9 (left hand side).

Moreover, biomimetic prosthesis should present rigidities along the isostatic lines that match those of the bone where it is placed.

Five regions of the prosthesis have been chosen to assign differentiated decreasing rigidities (right hand side of fig.9).

The state stress acting on the system and the observation of the isostatic lines described by Kummer (1986) has been used to define the morphology of the porosity characterizing the trabecular structure of the stem and head of the prosthesis in different zones.

These regions should be characterized by different rigidities that progressively decrease from the highly rigid head (region 1 in fig.9) down to the highly flexible (region 5 in fig.9).

The isostatic lines and the morphology differences of the isorigidity trabecular regions are better appreciated by comparing trabecular structure of isostatic and isorigidity regions (figures 10b and 10c).

The diverse orientation and shape of the Titanium trabecular porosity is evident in the internal structure reported in the fig. 10 (b) and (c).

An example of isorigidity trabecular structure obtained by Ti alloy powder sintering by Electron Beam is reported in figure 10d.

The apparent elastic moduli of the trabecular structures in the isorigidity regions vary from 20 GPa for region 1 to 11, 8, 4 and 1 GPa for regions 2, 3, 4 and 5, respectively.

The orientation of trabecular system observed in the lower left of figure 10b, which is observed in the area of the prosthesis stem corresponding to the femur epiphysis subjected to tensile stresses, would preserve and provide the requested continuity of the rigidity in each specific region reported in figure 10a.

The previously CT computed values of bone densities have been then related to the isostatic lines of cortical and trabecular bone of the sound and resected femur epiphysis by considering each tethraedric mesh element oriented along the stress isostatic directions observed in the femur proximal epiphysis. Analogously, the same criterion has been applied (isostatic line orientation) for the biomimetic trabecular Titanium prosthesis material definition.

FEM analysis: sound and prothesized femur, with traditional rigid and new flexible Ti prostheses

On the basis of the 3D models of the sound femur and the resected femurs prosthesized with rigid and flexible stem prosthesis, a structural evaluation under the same loading conditions has been developed and the results critically compared.

The characteristic biometric parameters of the patient femur-hip system have been carried out before running the structural analysis.

The 3Matic software has been used to identify the direction of the loading axis, the center of the proximal epiphysis head sphere, and the center of the joint epicondyle and mechanical axis of rotation of the knee.

Namely, the following parameters have been measured: the mechanical axis between the centers of the proximal and distal epiphyses, the angle of 143.40 ° between the femoral neck and diaphysis axes, and the divergence of 36.65° between the neck axis with the axis of epicondyles.

Finally (before running the FEM structural analyses) the physiological loads and constrains have been evaluated.

Loading condition and imposed constrains

The equilibrium monopodalic posture has been considered as an extreme severe loading condition with a rotation around the center of hip joint where the gluteus muscular force equilibrates the moment imparted by the body weight force. Considering the weight of the patient of 100 Kg, the gluteus muscular force that is applied to the great trochanter is of 1800 N and the hip joint reaction force is 2740 N (these values have been calculated using the biometric data of figure 11).

The reaction force of the Gluteus muscle has been uniformly distributed around 100 nodes of great trochanter FEM model area, while reaction force in the hip joint has been spread over 50 nodes of the femur head (see upper left in figure 12).

 

  1. Results and discussion

 

 In this study we defined the procedure to prepapre a biofidel model the femur able to correctly describe its biomechanical structural behavior.

Von Mises strain criterion has been used to compare and validate the new designed trabecular flexible prostheses.

Von Mises is an energetic criterion that can properly quantify the ability of the bone to withstand high loads (Figure 12).

The different structural behavior of the two models is also makes evident by the Von Mises stress distribution, clearly different at the posterior and the front level (right hand end of fig.12 and figure 13a).

Posteriorly, strain distribution confirms the presence of bending effects, with the highest values homogeneously distributed in the anatomical diaphysis regions (figure 13a).

Figure 13 clearly shows the significant modification of the stress distribution in the sound femur (fig.13a), the resected femur with the rigid Ti prosthesis (fig.13b), and with the flexible stem prosthesis (fig.13c). As already specified in a previous paper, the rigidly prosthesized femur misses the physiological flexural stresses distribution of a sound diaphysis (fig.13a) leading to a stress concentration only in the medial region a “stress shielded” proximal  (due to the stem high rigidity that does not allow diaphysis physiological flexure).

The absence of stress (and related strains state) could induce, over the time, significant bone structure modification (bone reabsorption).

Conversely, the biomimetic flexible trabecular prosthesis model, realizes a more uniform strains distribution (fig.13c) that better mimics in terms of stress distribution and intensities the action of the load on a sound femur (fig.13a).

 

  1. Conclusions

 

 The innovative aspects of our work resides in the fact that, contrariwise to the currently used prosthesis that are only designed for the replacement of a damaged hip joint, the proposed biological hybrid prostheses are able to be completely biomimetic since mimicking biological stress distribution that  stimulates physiological tissues regeneration.

The average life of a prosthesis today is about 10/15 years, while the new "biomimetic prosthesis" will have a longer life that can be estimated to be over 20/25 years.

This aspect is very important because the average human lifespan is progressively increasing, and this consequently increases the number of orthopedic surgical operation, and social and health care related costs.

The design and fabrication of customized different densities trabecular porous structures of innovative “hip joint biomimetic prosthetic systems”, which could better integrate with the physiological biomechanics of the femur where they are implanted, is then possible by correctly using such biofidel models.

The aim of the investigation presented here is to create added value by combining the existing research on biomechanical results with innovative prosthesis design and structural simulation activities.

Transfer results across academic and industrial research on biomechanics and clinical studies lead to an acceleration of the innovation and creation of the profit while improving the quality of life of the patients with prostheses.

This work identifies a number of design criteria to foster the potential to enable new medical therapies contributing to customized health care, to create and improve the technological basis and to increase the resource efficiency in the context of industrial and manufacturing processes.

The new type of biomimetic implants can find applications in orthopedics of knee, ankle, hip, shoulder, and spine.

A further field of application of the product is surgical oncology to support and facilitate bone regeneration resulting in massive losses due to interventions of removal of both primitive and metastatic tumors.

The prosthetic system might allow a better functional recovery by promoting bone recreation to ensure a good hold of the load, even going to impact on the quality of life of the individual patient, already seriously compromised by the basic oncological pathology.

The concept of combining a metal support structure (to guarantee the resistance to the load) with a biomimetic scaffold (which promotes regeneration) finds application in all those sectors of the therapeutic surgery involving the removal of bone and requiring a stimulation of the regeneration of resected tissue.

 

  1. References

 

Annunziata M, Aversa R, Apicella A, Annunziata A, Apicella D, Buonaiuto C, Guida L (2006); In vitro biological response to a light-cured composite when used for cementation of composite inlays, Dental Materials Volume 22, Issue 12, Pages 1081-1085

Apicella D, Aversa R, Ferro E, Ianniello D,  Apicella A (2010); The importance of cortical bone orthotropicity, maximum stiffness direction and thickness on the reliability of mandible numerical models,  Journal of Biomedical Materials Research - Part B Applied Biomaterials, Volume 93, Issue 1, April, 150-163 (8): doi: 10.1002/jbm.b.31569

Apicella D, Veltri M, Balleri P, Apicella A and Ferrari M (2011); Influence of abutment material on the fracture strength and failure modes of abutment-fixture assemblies when loaded in a bio-faithful simulation, Clinical Oral Implants Research, Volume 22, Issue 2, February, 182-188: DOI: 10.1111/j.1600-0501.2010.01979.x

Apicella D, Aversa R, Tatullo M, Simeone M, Sayed S, Marrelli M, Apicella A (2015); Direct restoration modalities of fractured central maxillary incisors: A multi-levels validated finite elements analysis with in vivo strain measurements, Dental Materials Volume 31, Issue 12, 1 December, e289-e305, DOI: 10.1016/j.dental.2015.09.016

Ashman RB, Cowin SC, Van Buskirk WC, and Rice JC (1984); A continuous wave technique for the measurement of the elastic properties of cortical bone, J. Biomechanics Vol. 17 No. 5, 349-361, doi:10.1016/0021-9290(84)90029-0

Ashman RB, Rho JY, (1988); Elastic modulus of trabecular bone material. J. Biomechanics, vol 21, 177-181: doi:10.1016/0021-9290(88)90167-4

Aversa R, Apicella D, Perillo L, Sorrentino R, Zarone F, Ferrari F, Apicella A (2009); Non-linear elastic three-dimensional finite element analysis on the effect of endocrown material rigidity on alveolar bone remodeling process. Dental materials 2009; 25: 678–690: DOI: 10.1016/j.dental.2008.10.015

Aversa, Raffaella; Petrescu, Relly Victoria V.; Apicella, Antonio; Petrescu, Florian Ion T.; 2017a Nano-Diamond Hybrid Materials for Structural Biomedical Application, American Journal of Biochemistry and Biotechnology, 13(1).

Aversa, Raffaella; Petrescu, Relly Victoria; Akash, Bilal; Bucinell, Ronald B.; Corchado, Juan M.; Berto, Filippo; Mirsayar, MirMilad; Chen, Guanying; Li, Shuhui; Apicella, Antonio; Petrescu, Florian Ion Tiberiu; 2017b Kinematics and Forces to a New Model Forging Manipulator, American Journal of Applied Sciences 14(1):60-80.

Aversa, Raffaella; Petrescu, Relly Victoria; Apicella, Antonio; Petrescu, Florian Ion Tiberiu; Calautit, John Kaiser; Mirsayar, MirMilad; Bucinell, Ronald; Berto, Filippo; Akash, Bilal; 2017c Something about the V Engines Design, American Journal of Applied Sciences 14(1):34-52.

Aversa, Raffaella; Parcesepe, Daniela; Petrescu, Relly Victoria V.; Berto, Filippo; Chen, Guanying; Petrescu, Florian Ion T.; Tamburrino, Francesco; Apicella, Antonio; 2017d Processability of Bulk Metallic Glasses, American Journal of Applied Sciences 14(2):294-301.

Aversa, Raffaella; Petrescu, Relly Victoria V.; Akash, Bilal; Bucinell, Ronald B.; Corchado, Juan M.; Berto, Filippo; Mirsayar, MirMilad; Chen, Guanying; Li, Shuhui; Apicella, Antonio; Petrescu, Florian Ion T.; 2017e Something about the Balancing of Thermal Motors, American Journal of Engineering and Applied Sciences 10(1).

Aversa, R., F.I.T. Petrescu, R.V. Petrescu and A. Apicella, 2016a. Biomimetic FEA bone modeling for customized hybrid biological prostheses development. Am. J. Applied Sci., 13: 1060-1067. DOI: 10.3844/ajassp.2016.1060.1067

Aversa, R.; Parcesepe, D.; Petrescu, R.V.; Chen, G.; Petrescu, F.I.T.; Tamburrino, F.; Apicella, A. 2016b Glassy Amorphous Metal Injection Molded Induced Morphological Defects, Am. J. Applied Sci. 13(12):1476-1482.

Aversa, R.; Petrescu, R.V.; Petrescu, F.I.T.; Apicella, A.; 2016c Smart-Factory: Optimization and Process Control of Composite Centrifuged Pipes, Am. J. Applied Sci. 13(11):1330-1341.

Aversa, R.; Tamburrino, F.; Petrescu, R.V.; Petrescu, F.I.T.; Artur, M.; Chen, G.; Apicella, A.; 2016d Biomechanically Inspired Shape Memory Effect Machines Driven by Muscle like Acting NiTi Alloys, Am. J. Applied Sci. 13(11):1264-1271.

Aversa, R.; Buzea, E.M.; Petrescu, R.V.; Apicella, A.; Neacsa, M.; Petrescu, F.I.T.; 2016e Present a Mechatronic System Having Able to Determine the Concentration of Carotenoids, Am. J. of Eng. and Applied Sci. 9(4):1106-1111.

Aversa, R.; Petrescu, R.V.; Sorrentino, R.; Petrescu, F.I.T.; Apicella, A.; 2016f Hybrid Ceramo-Polymeric Nanocomposite for Biomimetic Scaffolds Design and Preparation, Am. J. of Eng. and Applied Sci. 9(4):1096-1105.

Aversa, R.; Perrotta, V.; Petrescu, R.V.; Misiano, C.; Petrescu, F.I.T.; Apicella, A.; 2016g From Structural Colors to Super-Hydrophobicity and Achromatic Transparent Protective Coatings: Ion Plating Plasma Assisted TiO2 and SiO2 Nano-Film Deposition, Am. J. of Eng. and Applied Sci. 9(4):1037-1045.

Aversa, R.; Petrescu, R.V.; Petrescu, F.I.T.; Apicella, A.; 2016h Biomimetic and Evolutionary Design Driven Innovation in Sustainable Products Development, Am. J. of Eng. and Applied Sci. 9(4):1027-1036.

Aversa, R., Petrescu, R.V., Apicella, A., and Petrescu, F.I.T., 2016i Mitochondria are Naturally Micro Robots - A review, Am. J. of Eng. and Applied Sci. 9(4):991-1002.

Aversa, R.; Petrescu, R.V.; Apicella, A.; Petrescu, F.I.T.; 2016j We are Addicted to Vitamins C and E-A Review, Am. J. of Eng. and Applied Sci. 9(4):1003-1018.

Aversa, R., Petrescu, R.V., Apicella, A., and Petrescu, F.I.T., 2016k Physiologic Human Fluids and Swelling Behavior of Hydrophilic Biocompatible Hybrid Ceramo-Polymeric Materials, Am. J. of Eng. and Applied Sci. 9(4):962-972.

Aversa, R.; Petrescu, R.V.; Apicella, A.; Petrescu, F.I.T.; 2016l One Can Slow Down the Aging through Antioxidants, Am. J. of Eng. and Applied Sci. 9(4):1112-1126.

Aversa, R.; Petrescu, R.V.; Apicella, A.; Petrescu, F.I.T.; 2016m About Homeopathy or jSimilia Similibus Curenturk, Am. J. of Eng. and Applied Sci. 9(4):1164-1172.

Aversa, R.; Petrescu, R.V.; Apicella, A.; Petrescu, F.I.T.; 2016n The Basic Elements of Life's, Am. J. of Eng. and Applied Sci. 9(4):1189-1197.

Aversa, R.; Petrescu, F.I.T.; Petrescu, R.V.; Apicella, A.; 2016o Flexible Stem Trabecular Prostheses, Am. J. of Eng. and Applied Sci. 9(4):1213-1221.

Beaupre G.S. and Hayes W.C. (1985): Finite Element Analysis of a three dimensional open-celled model for trabecular bone. J. Biomech. Eng. Vol 107, pp 249-56, PMID: 4046566

Burnstein A., Reilly D. T., Martens M. (1976), Aging of bone tissue: Mechanical properties., J. of Bone and joint Surgery, vol. 58, pp. 82-86, https://www.researchgate.net/publication/21906817_Aging_of_Bone_Tissue_Mechanical_Properties

Carter D. R., Hayes W. C. (1977) The compressive behavior of bone as a two phase porous structure. J. of Bone and joint Surgery, 59A: 954, PMID: 561786

Dalstyra M., Huiskes R., Odgaard A. and Van Erning L. (1993) Mechanical and textural properties of Pelvic Trabecular Bone. J. Biomechanics Vol. 26 No. 4-5  pp. 349-361, DOI: 10.1016/0021-9290(93)90014-6

Frost HM. (1994), Wolff’s law and bone’s structural adaptations to mechanical usage: an overview for clinicians. Angle Orthod; 64:175–88.

Gramanzini M, Gargiulo, S Zarone F, Megna, R, Apicella A, Aversa, R, Salvatore M, Mancini M, Sorrentino R, , , Brunetti A (2016), Combined microcomputed tomography, biomechanical and histomorphometric analysis of the peri-implant bone: A pilot study in minipig model  Dental Materials, Volume 32, Issue 6, 1 June 2016, Pages 794-806: DOI: 10.1016/j.dental.2016.03.025

Huiskes R, Weinans H, Grootenboer HJ, Dalstra M, Fudula B, Slooff TJ. Adaptive bone remodeling theory applied to prosthetic-design analysis. J Biomech 1987;20:1135–1150.

Oh I., e Harris W H. (1976) Proximal distribution in the loaded femur. J. of Bone and Joint Surgery, vol. 60-A, no 1: PMID: 624762

Kummer B. (1986), Biomechanical principles of the statistics of the hip joint. A critical appraisal of a new theory, Zeitschrift fur Orthopadie und Ihre Grenzgebiete, Volume 124, Issue 2, 1986, Pages 179-187

Mirsayar, M. M., & Park, P. (2016). Modified maximum tangential stress criterion for fracture behavior of zirconia/veneer interfaces. Journal of the mechanical behavior of biomedical materials, 59, 236-240.

Mirsayar, M.M., Joneidi, V.A., Petrescu, R.V.V., Petrescu, F.I.T., Berto, F., 2017 Extended MTSN criterion for fracture analysis of soda lime glass, Engineering Fracture Mechanics 178:50–59, ISSN: 0013-7944, http://doi.org/10.1016/j.engfracmech.2017.04.018

Mullender MG, Huiskes R. (1995) A proposal for the regulatory mechanism of Wolff’s law. J Orthop Res;13:503–512. DOI: 10.1002/jor.1100130405

Perillo L, Sorrentino R, Apicella D, Quaranta A, Gherlone, E, Ferrari M, Aversa R, Apicella A (2010); Nonlinear visco-elastic finite element analysis of porcelain veneers: a submodelling approach to strain and stress distributions in adhesive and resin cement. The journal of adhesive dentistry, Volume 12, Issue 5, October 2010, Pages 403-413: ISSN: 14615185

Reilly D, and Burstein AH (1974); The mechanical properties of cortical bone. The J. Of bone and Joint Surgery, Vol. 56 A, No. 5 July, 1001-1021

Rohlmann A, Mossner U, Bergmann G and Kolbel R (1982); Finite Element Analysis and experimental investigations of stresses in a femur.  J. Biomed. Eng. Vol. 4: doi:10.1016/0141-5425(82)90009-7

Reilly DT and Burnestain A H (1975); The elastic and ultimate properties of compact bone tissue.  J. Biomechanics vol 8, pp 393-405, doi:10.1016/0021-9290(75)90075-5

Schiraldi C, D’Agostino A, Oliva A, Flamma F, De Rosa A, Apicella A, Aversa A, De Rosa M, (2004); Development of hybrid materials based on hydroxyethylmethacrylate as supports for improving cell adhesion and proliferation, Biomaterials, Vol. 25 17, 3645-3653: http://dx.doi.org/10.1016/j.biomaterials.2003.10.059

Sorrentino, R. Apicella, D.,  Riccio, C., Gherlone, E.d,  Zarone, F.,  Aversa, R.e,  Garcia-Godoy, F,  Ferrari M, Apicella A (2009); Nonlinear visco-elastic finite element analysis of different porcelain veneers configuration , Journal of Biomedical Materials Research - Part B Applied Biomaterials, Vol 91, Issue 2, 727-736; DOI: 10.1002/jbm.b.31449  (10)

Sorrentino R, Aversa R, Ferro V, Auriemma T, Zarone F, Ferrari M, Apicella A. (2007); Three-dimensional finite element analysis of strain and stress distributions in endodontically treated maxillary central incisors restored with different post, core and crown materials. Dent Mater; 23: 983–93: DOI: 10.1016/j.dental.2006.08.006

Taylor D, Hazenberg JG, Lee TC. (2007) Living with cracks: Damage and repair in human bone. Nat Mater;6:263–268.

Tamar Gottesman and Zvi Hashin, (1980), Analysis of viscoelastic behaviour of bones on the basis of microstructure. J. Biomechanics, Vol13 pp. 89-96: DOI: http://dx.doi.org/10.1016/0021-9290(80)90182-7

Weinans H, Huiskes R, Grootenboer HJ. (1992) The behavior of adaptive bone remodeling simulation models. J Biomech; 25:1425–1441. PMID: 1491020

 